| <p>On 23/02/2018 at 13:49 we started the DAQ in list mode.</p>
<p>After starting the DAQ we sent the 3.45MeV 7Li beam onto the CH2 930ug/cm2 target III.</p>
<p>The trigger rate ~ 2kHz, event multiplicity = 6.1, dead time ~ 9%.</p>
<p>The DAQ save file is: /Remote_Objs/RData_1/180223135345</p>
<p>Checking the ADC rate histograms:</p>
<ul>
<li>ADC 2 Channels 0->7 Rate ~ 1e3 (channels 0 and 5 still disabled)</li>
<li>ADC 2 Channel 8 dead</li>
<li>ADC 2 Channels 9->31 Rate ~ 1e4 - 1e5</li>
<li>ADC 4 Channels 0->15 Rate ~ 1e3</li>
<li>ADC 4 Channels 16 -> 29 and 31 Rate ~ 1e4 - 1e5</li>
<li>ADC 4 Channel 30 dead</li>
<li>ADC 6 Channel 31 Rate ~ 8e5</li>
</ul>
<p>Picking an S2 backstrip at random we found a proton detection rate ~ 100Hz, no high rate beam scatters were seen, as expected from the thick target. We decided to run for several hours.</p>
<p>We performed occasional checks during the run:</p>
<table border="1" cellpadding="1" cellspacing="1" style="width:500px">
<tbody>
<tr>
<td>Runtime (hh:mm:ss)</td>
<td>ADC # / Ch #</td>
<td>Area in region</td>
<td>Ratio to pin diode (ADC 6)</td>
</tr>
<tr>
<td colspan="1" rowspan="3">01:01:00</td>
<td>2/20</td>
<td>183850</td>
<td>61.47</td>
</tr>
<tr>
<td>6/31</td>
<td>2991</td>
<td> </td>
</tr>
<tr>
<td>4/24</td>
<td>262051</td>
<td>87.61</td>
</tr>
<tr>
<td colspan="1" rowspan="3">01:30:37</td>
<td>2/20</td>
<td>269255</td>
<td>61.49</td>
</tr>
<tr>
<td>6/31</td>
<td>4380</td>
<td> </td>
</tr>
<tr>
<td>4/24</td>
<td>377251</td>
<td>82.94</td>
</tr>
<tr>
<td colspan="1" rowspan="3">02:03:22</td>
<td>2/20</td>
<td>362581</td>
<td>58.07</td>
</tr>
<tr>
<td>6/31</td>
<td>5987</td>
<td> </td>
</tr>
<tr>
<td>4/24</td>
<td>501628</td>
<td>77.40</td>
</tr>
<tr>
<td colspan="1" rowspan="3">02:30:34</td>
<td>2/20</td>
<td>437224</td>
<td>57.86</td>
</tr>
<tr>
<td>6/31</td>
<td>7277</td>
<td> </td>
</tr>
<tr>
<td>4/24</td>
<td>600118</td>
<td>76.35</td>
</tr>
<tr>
<td colspan="1" rowspan="3">03:19:02</td>
<td>2/20</td>
<td>561345</td>
<td>53.80</td>
</tr>
<tr>
<td>6/31</td>
<td>9584</td>
<td> </td>
</tr>
<tr>
<td>4/24</td>
<td>765584</td>
<td>
<p>71.72</p>
</td>
</tr>
</tbody>
</table>
<p>Where the area regions used (in channels) for each ADC Channel are:</p>
<p>ADC 2 Ch 20: 0->552</p>
<p>ADC 6 Ch 31: 1302->1475</p>
<p>ADC 4 Ch 24: 20 ->549</p>
<p>The ratios use the number of counts since the previous check. The ratios are decreasing as expected, because this indicates that fewer protons are being detected during the irradiation time.</p>
<p>At 17:08 the DAQ was stopped at runtime 03:19:02</p>
<p>The DAQ monitor dead time ~ 5%. Event multiplicity ~ 6.0, # Good events = 23917715.</p>
<p>We checked the beam current before the chamber, FC-E0 ~ 190pA.</p>
<p> </p>
<p>See attached "180223_S2FrontStrip_ADC2CH20.png" for sample front S2 strip spectrum</p>
<p>See attached "180223_S2BackStrip_ADC4CH24.png" for sample back S2 strip spectrum</p>
<p>See attached "180223_PinDiode_ADC6CH31.png" for sample pin diode spectrum</p> |