<p>Date: 07/04/2018</p>
<p>FC - 4 = 410 - 260 = 150 pA</p>
<p>We moved the target ladder to position V (35.022mm) - CH2 Target IV</p>
<p>At 15:09 we started the DAQ for 5.0 MeV 7Be in 2+ charge state bombarding CH2 IV</p>
<ul>
<li>DT = 0%</li>
<li>Trigger rate ~ 10Hz</li>
<li>Event mult = 69.7</li>
<li>Save name recorded on paper logbook as 180407111531 -> same as save name used for 7Li -> emailed Raffaele to check whether there are other saves starting with 180407......</li>
</ul>
<p>At 16:39 the DAQ was stopped, RT = 01:29:07, # good events = 68068</p>
<p>Checked beam current due to low trigger rate, FC-4 = 405 - 260 = 155 pA</p>
<p>Central and downstream irises closed, target ladder set to position VII - 3mm collimator</p>
<p>FC - E1 = 63 - 5 = 58 pA, Transmission ~ 33%</p>
<p>Stopped beam, Time ~ 17:00, decided to stop and dismount detectors</p>
<p>17:00 pressures read:</p>
<ul>
<li>TPU1 = 2.2e-6 mBar</li>
<li>TPU2 = 8.2e-7 mBar</li>
<li>TPD1 = 1.3e-6 mBar</li>
<li>TPD2 = 1.5e-6 mBar</li>
<li>CHUP = 1.0e-5 mBar</li>
<li>CHDW = 2.7e-5 mBar</li>
</ul>
<p>Detector status:</p>
<ul>
<li>Central S2 bias = -129.9V, leak I = 18.71uA</li>
<li>diode bias = -70.1V, leak I = 0.00uA</li>
<li>Downstream S2 bias = -130.1V, leak I = 2.84uA</li>
</ul> |