Entry time:
Wed Nov 21 08:53:06 2018
Author
*
:
Subject
*
:
<p>On 19/11/2018:</p> <p>During the vacuum test - see eLog entry #43 we performed noise tests on the preamplifiers with no S2 detector connected</p> <p>First we performed a "tabletop" test - the preamplifiers were not connected to the chamber.</p> <p>Preamp I for the near S2 is connected to RAL modules (reading left to right): X, XII, XIII, R, S, Q, U, and W, via long (~5m) preamp - ICU cables 13, 14, 15, and 16</p> <p>Preamp II for the far S2 is connected to RAL modules (reading left to right): I, II, XI, IV, V, VI, VII, VIII, via long (~5m) preamp - ICU cables 0, 1, 2, and 3</p> <p>We powered preamps I and II with + and - 15V. The power supply shows the correct currents drawn: ~1.5A for +ve and ~0.75A for -ve</p> <p>We found our test 16 pin cables #45 had a dead channel 0, we switched to cable #46 which shows all 8 channels working.</p> <p>We found the following dead channels on the RAL modules:</p> <table border="1" cellpadding="1" cellspacing="1" style="width:500px"> <tbody> <tr> <td>RAL</td> <td>CH #</td> </tr> <tr> <td>U</td> <td>0</td> </tr> <tr> <td>W</td> <td>0, 7</td> </tr> <tr> <td>I</td> <td>0, 5 (Both disabled in ICU)</td> </tr> <tr> <td>II</td> <td> <p>0, 2</p> </td> </tr> </tbody> </table> <p>We mounted preamps I and II to the top of the chamber. We mounted preamp I on the far S2 mount and preamp II on the near S2 mount, however we were not connecting to the S2's and so continued with the electronics tests.</p> <p>Now we saw the following dead channels on the RAL modules:</p> <table border="1" cellpadding="1" cellspacing="1" style="width:500px"> <tbody> <tr> <td>RAL</td> <td>CH #</td> </tr> <tr> <td>W</td> <td>7</td> </tr> <tr> <td>I</td> <td>0, 5 (disabled in ICU)</td> </tr> <tr> <td>II</td> <td>0, 2</td> </tr> </tbody> </table> <p>We checked the signal to noise ratio for RAL X CH0:</p> <p>The input pulser amplitude = 50mV, the RAL analogue amplitude = 950 mV, the RMS noise = 40mV. The S/N ~4.2%.</p> <p>We stopped tests for the day.</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown