AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
CIRCE
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Sat Nov 24 16:29:10 2018
Author
*
:
Subject
*
:
<p>On 23/11/2018:</p> <p>10:15 started pumping down chamber with Ecodry 2</p> <p>RAL thresholds:</p> <ul> <li>First 4 RALs = 10mV</li> <li>5th RAL = 17mV</li> <li>6th RAL = 20mV</li> <li>7th and 8th RALs ECL disconnected</li> <li>9th - 16th RALs not in use (no near S2)</li> </ul> <p>10:30 pressures ~ 1mBar, opened T1600 backing valve, Turbos switched on</p> <p>10:35 pressures (mBar):</p> <ul> <li>TPU1 = 5.2e-5</li> <li>TPU2 = 6.4e-6</li> <li>TPD1 = 6.0e-6</li> <li>TPD2 = 8.7e-6</li> <li>CHUP = 2.3e-5</li> <li>CHDW = 5.5e-5</li> </ul> <p>Biased detectors: Far S2 bias = -130V, leak I = 1.54uA. Diode bias = -70V, leak I = 0.02uA.</p> <p>Replaced diode TFA, now using channel 4 on new TFA. Diode connected to 3rd ADC channel 16 (counting from 0)</p> <p>Started pulser only run</p> <ul> <li>Run # 181123111015</li> </ul> <p>Closed far S2 iris (82.5mm actuator setting). Set target ladder to blank 10mm target frame.</p> <p>Disconnected vacuum gauges located below chamber on same side as target ladder is mounted - noise improved</p> <p>Started pulser only run</p> <ul> <li>Run # 181123113237</li> </ul> <p> </p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown