| <p>On 23/11/2018:</p>
<p>10:15 started pumping down chamber with Ecodry 2</p>
<p>RAL thresholds:</p>
<ul>
<li>First 4 RALs = 10mV</li>
<li>5th RAL = 17mV</li>
<li>6th RAL = 20mV</li>
<li>7th and 8th RALs ECL disconnected</li>
<li>9th - 16th RALs not in use (no near S2)</li>
</ul>
<p>10:30 pressures ~ 1mBar, opened T1600 backing valve, Turbos switched on</p>
<p>10:35 pressures (mBar):</p>
<ul>
<li>TPU1 = 5.2e-5</li>
<li>TPU2 = 6.4e-6</li>
<li>TPD1 = 6.0e-6</li>
<li>TPD2 = 8.7e-6</li>
<li>CHUP = 2.3e-5</li>
<li>CHDW = 5.5e-5</li>
</ul>
<p>Biased detectors: Far S2 bias = -130V, leak I = 1.54uA. Diode bias = -70V, leak I = 0.02uA.</p>
<p>Replaced diode TFA, now using channel 4 on new TFA. Diode connected to 3rd ADC channel 16 (counting from 0)</p>
<p>Started pulser only run</p>
<ul>
<li>Run # 181123111015</li>
</ul>
<p>Closed far S2 iris (82.5mm actuator setting). Set target ladder to blank 10mm target frame.</p>
<p>Disconnected vacuum gauges located below chamber on same side as target ladder is mounted - noise improved</p>
<p>Started pulser only run</p>
<ul>
<li>Run # 181123113237</li>
</ul>
<p> </p> |