On 28/11/2018:

Performed pulser walkthrough, Run # 181128090835

  • Run time: 1986s
  • Pulser ON, amplitude changed
  • Target: None
  • Beam: None
  • DT ~ 11%
  • See attached picture 1 for sample pulser walkthrough spectrum, ADC 2 channel 3 (ADC 3 in DAQ file browser)

Pulser amplitudes swept through (measured amplitude checked after collecting pulser walkthrough to ensure no additional load was placed on pulser):

Pulser Amplitude Setting Pulser Measured Amplitude (mV)
70000 366
60000 318
50000 268
40000 206
30000 155
20000 102
18000 92.4
16000 82.8
14000 72.8
12000 62.4
10000 51.8
8000 42.0
6000 32.0
4000 22.0
2000 11.4

After walkthough pulser amplitude setting set back to 70000, all other settings left unchanged

Far S2 and diode biases switched off. Preamp power off. Line valve and FCC valve closed. Turbos off. Manual and T1600 backing valves closed. Ecodry 1 and 2 off. Waited for turbos to spin down.

By 10:20 turbos had finished spinning, vented chamber ready to mount alpha source.

Attachment 1: 182811_PulserWalkthrough.jpg  75 kB  | Hide | Hide all