<p>On 28/11/2018:</p>
<p>Mounted 12um mylar in front of far S2 - see attached pictures 1 and 2</p>
<p>In air: Far S2 biased to -130.0V, leak I = 1.61uA. Diode biased to -70.1V, leak I = 0.01uA -> 0.00uA. Rate meter ~ 4500Hz.</p>
<p>Detectors switched off</p>
<p>17:50 started Ecodry 1 and 2, started opening manual valve.</p>
<p>18:00 Manual valve fully opened, T1600 backing valve opened, Turbos switched on.</p>
<p>Once pressures ~ 1e-5mBar, far S2 bias = -130.0V, leak I = 1.59uA. Diode bias = -70.1V, leak I = 0.00uA.</p>
<p>18:08 Pressures (mBar):</p>
<ul>
<li>TPU1 = 3.0e-5</li>
<li>TPU2 = 6.0e-5</li>
<li>TPD1 = 5.2e-6</li>
<li>TPD2 = 8.7e-6</li>
<li>CHUP = Sensor Off</li>
<li>CHDW = No Sensor</li>
</ul>
<p>Run # 181128181421</p>
<ul>
<li>Run time: 1799s</li>
<li>Pulser ON, as before</li>
<li>Alpha source</li>
<li>S2 mylar: 12um</li>
<li>DT ~ 12%</li>
</ul>
<p>Started overnight run:</p>
<p>Run # 181128184505</p>
<ul>
<li>Run time: 51706s</li>
<li>Pulser ON, as before</li>
<li>Alpha source</li>
<li>S2 mylar: 12um</li>
<li>DT ~ 8%</li>
<li>See attached picture 3 for sample far S2 front strip spectrum</li>
</ul>
<p>18:50 End of day check:</p>
<p>Far S2 bias = -130.0V, leak I = 1.60uA.</p>
<p>Diode bias = -70.1V, leak I = 0.00uA.</p>
<p>Pressures (mBar):</p>
<ul>
<li>TPU1 = 3.8e-6</li>
<li>TPU2 = 1.2e-6</li>
<li>TPD1 = 1.4e-6</li>
<li>TPD2 = 1.6e-6</li>
<li>CHUP = Sensor Off</li>
<li>CHDW = No Sensor</li>
</ul>
<p>On 29/11/2018: Alpha run stopped</p>
<p>9:05: Far S2 bias = -130.0V, leak I = 1.70uA. Diode bias = -70.1V, leak I = 0.00uA.</p>
<p>Detector biases and preamps off.</p>
<p>Pressures (mBar):</p>
<ul>
<li>TPU1 = 7.6e-7</li>
<li>TPU2 = 3.5e-7</li>
<li>TPD1 = 7.7e-7</li>
<li>TPD2 = 7.3e-7</li>
<li>CHUP = Sensor Off</li>
<li>CHDW = No Sensor</li>
</ul>
<p>Closed manual and T1600 backing valves. Stopped turbos. Stopped Ecodry 1 and 2.</p>
<p>Switched off Nim bin, RAL crate, and CAMAC crate</p>
<p>9:45 Opened venting valve</p> |