AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
|
DESPEC |
 |
|
Message ID: 109
Entry time: Wed Dec 18 17:39:27 2019
|
Author: |
TD |
Subject: |
Wednesday 18 December |
|
|
18.40 Detector biases & leakage currents OK - attachment 1
FEE64 temperatures OK - attachment 2
FEE64 good event statistics OK - attachment 3
3x <20k, 8x <50k, 11x <100k
note - all 12x FEE64s zero disc-info #6
system wide checks - see attachments 4-9
master clock status fail
WR decoder status - aida08 fails
18.49 GSI WR status control & timestamps - attachments 10 & 11
18.50 check ASIC control *all* FEE64s *all* ASICs
18.58 Rate, Stat, 1.8.L and 1.8.H spectra all FEE64s - attachments 12-17
OK except aida10 no HEC data & low amplitude LEC data
19.00 *all* histograms zero'd
19.03 NewMerger, TapeServer and full screen - attachments 18-20
19.07 Options, ASIC settings aida09 & aida10 - attachments 21-23
20.15 collect all WR status erro counters for baseline
WR decoder status - no errors
21.56 FRS DAQ crashed and not currently recoverable
Stop current run and run background alphas overnight
21.57 run stopped file 171219/R1_304
slow comparator threshold 0xf -> 0x64
check ASIC control *all* FEE64s *all* ASICs
BNC PB-5 attenuation x1 -> x1000
*all* histograms zero'd
22.02 DAQ start file 171219/R2
background alphas
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|