Entry  Wed Jan 15 14:53:41 2020, NH, AIDA Noise 14.01.2019 6x
    Reply  Mon Jan 20 14:01:23 2020, NH, AIDA Noise 20.01.2019 6x
Message ID: 120     Entry time: Wed Jan 15 14:53:41 2020     Reply to this: 121
Author: NH 
Subject: AIDA Noise 14.01.2019 
Checking noise

DESPEC crate entirely offline except:
VME Crate (for VETAR2)
AIDA (All)

Figs 1-2: Waveforms (Odd/Even)
Figs 3-4: Zoomed In (Odd/Even)
Fig 5: Good Event Stats
Fig 6: Merger Rate

Noise still present, despite nothing on DESPEC.

Things to check: 
Turn off VME (kills time stamps but might be OK for waveform test?)
Check floor PSU under AIDA is offline too 

Not sure that other FRS systems (e.g. Ion Catcher) is offline - I think RF may be on.