Author
*
:
Subject
*
:
> aida10 has no ASIC events when running without pulser > > DSSD ribbon cable was replaced, no change > > 29.01.2020 - change AIDA half adapter board, no change > > Swapped HV (braid) from aida10 to aida12 - no change in leakage current, now testing data... > Looks the same (no data) fig 2 > > Very hot channel in aida08 as well: > - Channel 104 in histogram Stat = 4.7.L fig 1 > > Change FEE64 card? > Any other checks possible? > > 31.01.2020: > > Installed new FEE64 card aida10: d8:80:39:41:ba:89 > Updated dhcpd.conf > System boots OK. > returned bias ground to aida10 for symmetry reasons > Firmware updated from 0x12110103 to 0x18430701 > > Left recording alpha data over the weekend: > See stats in fig 3 > > DSSD 1 and 2 look excellent (one broken strip in DSSd 1)? > > aida10 still no counts? > aida09 looks very... weird as well? > aida11 and aida12 seem ok? > > Detector fault?