Entry  Wed Jan 29 12:22:13 2020, NH, aida10 troubleshooting aida08-noise.pngaida10-noitems.pngStats.png
    Reply  Wed Jan 29 19:09:51 2020, NH, aida10 troubleshooting 
Message ID: 124     Entry time: Wed Jan 29 19:09:51 2020     In reply to: 122
Author: NH 
Subject: aida10 troubleshooting 
> aida10 has no ASIC events when running without pulser
> 
> DSSD ribbon cable was replaced, no change
> 
> 29.01.2020 - change AIDA half adapter board, no change
> 
> Swapped HV (braid) from aida10 to aida12 - no change in leakage current, now testing data...
>  Looks the same (no data) fig 2
> 
> Very hot channel in aida08 as well:
>  - Channel 104 in histogram Stat = 4.7.L  fig 1
> 
> Change FEE64 card?
> Any other checks possible?

when running the pulser in 10 do all the channels respond correctly ?
what happens if you drop the shaper threshold into the noise ?