DSSD#3 (most downstream) was replaced Noted that old DSSD wafer was not sitting properly - detached - in PCB, PCB flexing and/or inadequate epoxy gluing New DSSD#3 - 2998-22 - 1006 um - 100 V - ~4 uA Old Detector removed was 3208-14 Fig 1. HV Control Fig 2. Temps Fig 3. Stats Fig 4. Rates Fig 5-6: Waves Fig 7-8: Pulser Peaks Start alpha run at 17:48, 20FEB20, R1 Slow comparator 0x64 Pulser off Fig 9 : Alpha stats Fig 10: Merger stats Fig 11: Merger main page Fig 12: Timestamp errors in merger console
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
hv.png
Original size: 502x344
Attachment 2:
temps.png
Original size: 1920x1080
Attachment 3:
stats.png
Original size: 1920x1080
Attachment 4:
rates.png
Original size: 1920x1080
Attachment 5:
waves_even.png
Original size: 1920x1080
Attachment 6:
waves_odd.png
Original size: 1920x1080
Attachment 7:
pulser_even.png
Original size: 1920x1080
Attachment 8:
pulser_odd.png
Original size: 1920x1080
Attachment 9:
alpha_stats.png
Original size: 1920x1080
Attachment 10:
merger.png
Original size: 1418x1028
Attachment 11:
merger_stats.png
Original size: 1418x1028
Attachment 12:
merger_errors.png
Original size: 709x465
Attachment 13:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown