Looking at the HV bias in a scope to see if it is the source of noise especially on the p+n side Channel 1: HV Core channel 4 Channel 2: HV Braid Channel 4 Math: Ch1 - Ch2 100V bias (100 uA leakage current for 1 MOhm) No obvious frequency components or large frequency components? I do not know how this compares to RIKEN or STFC-DL Updated to put the right frequency range (misread before) Fig 3-4: Normalish looking data Fig 5-6: Examples of noise appearing randomly on math line Fig 7: 1ms long trace, noise stopped appearing Update 26.02.2020 Fig 8: Trace with all 4 HV channels off Fig 9: Trace with only oscilloscope biased Fig 10: Trace with braids unplugged from all 3 FEE64s (only channel 4 connected) Things to try: Joining clean earth & chassis ground? Checking HV jumper? Finding 'scope probes and testing NIM socket?
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
SCRN0057.PNG
Original size: 640x480
Attachment 2:
SCRN0054.PNG
Original size: 640x480
Attachment 3:
SCRN0059.PNG
Original size: 640x480
Attachment 4:
SCRN0060.PNG
Original size: 640x480
Attachment 5:
SCRN0071.PNG
Original size: 640x480
Attachment 6:
SCRN0073.PNG
Original size: 640x480
Attachment 7:
SCRN0075.PNG
Original size: 640x480
Attachment 8:
SCRN0078.PNG
Original size: 640x480
Attachment 9:
SCRN0079.PNG
Original size: 640x480
Attachment 10:
SCRN0080.PNG
Original size: 640x480
Attachment 11:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown