AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 15     Entry time: Wed Jan 23 08:12:01 2019
Author: CA, TD, NH, VP 
Subject: January 22nd 2019 
Attachment 1 - 1.8.W waveform spectra with pulser signal

Attachment 2 - 1.8.L pulser peak spectra

widths: aida01 - 16.06
       
        aida04 - 20.68

        aida05 - 14.03

        aida08 - 73.02

        aida09 - 669.29

        aida11 - 405.35

        aida12 - 397.6

Attachment 3 - good event statistics
               fast comparator threshold (LEC) at 0xff
               FEEs 8,9,11,12 running hot

Attachment 4 - FEE64 temperatures - all ok

System wide checks ok, except aida02 fails ADC calibration

Leak found in seal on cooling pipe, switched power off