AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 172     Entry time: Fri Mar 5 22:53:05 2021
Author: OH 
Subject: Thursday 5th of March 
There has been no beam for much of the day. It was lost at 13:00 and not regained until 22:00 German time.
At some point during the day the White Rabbit timestamp cable was unplugged from AIDA. This caused it lose correlation with the rest of the system.
At first a reset and setup was performed to try and regain correlation but that did not prove successful.
A full powercycle was then performed at ~22:30 German time which upon restarting the DAQ solved the issue.

The plan is to overnight run some fragments, for the purposes of observing the isomers. This data is not essential and as such we will not be writing to disk.

Beam plug is in front of AIDA. This is causing a huge spike in the Ge.
ELOG V3.1.4-unknown