10.30 AIDA setup and DAQ start complete writing to file TapeData/NULL/R1 FEE64 Temperatures and Good event statistics ok (attachments 1 and 2) 11.00 results of analyser - shows significant number of timewarps in aida07, and a few in aida01-04 11.24 compared timestamps between mbs and AIDA, agree to 8 digits (attachment 3) 12.49 DAQ and Merger stopped and restarted Raised slow comparator threshold from 0xa to 0x64 Raised fast comparator threshold (LEC/MEC) from 0xf to 0xff 12.51 Good event statistics: aida08 running at a normal rate pulser peak (1.8.L) no longer has noise shoulder (attachment 4) writing to file TapeData/NULL/R2 13.58 stopped writing to file. Pulser frequency increased to 1 kHz writing to file TapeData/NULL/R3 15.00 file R4 - DAQ startup 17.58 trimmed and re-crimped ground wire on aida08 lowered pulser freq back to 50Hz reduced slow comparator threshold to 0xa, fast comparator threshold (LEC) to 0xf good event stats ok, aida08 no longer running hot (attachment 6) 1.8.L and 1.8.W spectra of aida08 (attachment 5) pulser peak width now at 55.71 FEE64 temperatures ok (attachment 7)
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
240119_stats.png
Original size: 1273x408
Attachment 2:
240119_temp.png
Original size: 1276x410
Attachment 3:
240119_timestamps.png
Original size: 1279x992
Attachment 4:
240119_fee8_18L.png
Original size: 1279x969
Attachment 5:
240119_spectra.png
Original size: 1277x764
Attachment 6:
1740Statsafter240119.png
Original size: 1279x407
Attachment 7:
240119_tempafter.png
Original size: 1275x408
Attachment 8:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown