AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 21     Entry time: Thu Jan 24 14:53:26 2019
Author: TD 
Subject: Analyses GSIJan19/R2_0 - R4_0 
Attachment 1 - summary analysis R2_0

               increased slow comparator & LEC/MEC fast comparators cf. R1_0 to minimise random noise

Attachment 2 - summary analysis R3_0

               per R2_0 but BNC PB-5 pulser frequency increased from c. 2Hz to c. 1kHz
               pulser to *all* channels 12x FEE64s

Attachment 3 - summary analysis R4_0

               open TapeSever file, DAQ GO, Start ASIC Readout (*correct* DAQ start sequence)

Observations

- some RESUME but *zero* PAUSE data items observed

- information types 4 & 5 not equal for R2_0 and almost, but not quite, equal for R3_0 & R4_0 

- information types 4 & 5 start data file with 'correct' DAQ start sequence

           
Attachment 1: WR_sample_R2_0_24Jan19  26 kB  | Show | Show all
Attachment 2: WR_sample_R3_0_24Jan19  51 kB  | Show | Show all
Attachment 3: WR_sample_R4_0_24Jan19  11 kB  | Show | Show all
Attachment 4: WR_sample_R4_0_24Jan19.log  31 kB  | Show | Show all
ELOG V3.1.4-unknown