Attachment 1 - summary analysis R2_0
increased slow comparator & LEC/MEC fast comparators cf. R1_0 to minimise random noise
Attachment 2 - summary analysis R3_0
per R2_0 but BNC PB-5 pulser frequency increased from c. 2Hz to c. 1kHz
pulser to *all* channels 12x FEE64s
Attachment 3 - summary analysis R4_0
open TapeSever file, DAQ GO, Start ASIC Readout (*correct* DAQ start sequence)
Observations
- some RESUME but *zero* PAUSE data items observed
- information types 4 & 5 not equal for R2_0 and almost, but not quite, equal for R3_0 & R4_0
- information types 4 & 5 start data file with 'correct' DAQ start sequence
|