AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Message ID: 220     Entry time: Wed Apr 14 12:14:47 2021
Author: PJCS 
Subject: Corrections and changes 

14/4/21 @12:00 UK time

Corrected the failure of FADC re-calibration All modules. The problem was a missing > in the .tml file

Edited the sys.tcl file in /MIDAS/TclHttpd/Html/RunControl to comment out where the status of the waveform enable was able to stop the creation of the .W spectra. This will now function at SETUP.

Edited the sys.tcl file to remove the test for Master module at the enable of the Correlation Scalar readout. All modules will now be enabled.

Successfully re-calibrated the LMK3200 in aida09 to remove the clock fault reported in System Wide Checks.

Successfully "sync ASIC clocks" followed by calibrate all module FADCs ( except aida09 which required seperate attention. )

All the Edits first carried out and tested at Daresbury T9 system.

ELOG V3.1.4-unknown