Entry time:
Tue May 4 13:53:22 2021
Author
*
:
Subject
*
:
NH has re-checked alignment of AIDA adaptor PCBs cf. Elog:290 14.40 Rate spectra - attachment 1 adc data items - attachment 2 slow comparator 0xa stats for p+n much improved, n+n remain poor 1.8.W spectra - attachments 3 & 4 1.8.L spectra - attachments 5 & 6 pulser peak widths aida01 69 ch FWHM, aida02 247 ch FWHM 17:31 Bias lowered to 100V on DSSD2 after noticing the leakage current has increased 2uA over the previous 3 hours. - attachment 7
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
Screenshot_2021-05-04_Spectrum_Browser_aidas-gsi.png
Original size: 2385x1163
Attachment 2:
Screenshot_2021-05-04_Statistics_aidas-gsi.png
Original size: 2144x1035
Attachment 3:
Screenshot_2021-05-04_Spectrum_Browser_aidas-gsi(1).png
Original size: 2131x1138
Attachment 4:
Screenshot_2021-05-04_Spectrum_Browser_aidas-gsi(2).png
Original size: 2131x1136
Attachment 5:
Screenshot_2021-05-04_Spectrum_Browser_aidas-gsi(3).png
Original size: 2131x1138
Attachment 6:
Screenshot_2021-05-04_Spectrum_Browser_aidas-gsi(4).png
Original size: 2131x1134
Attachment 7:
210504_Leakage.png
Original size: 1333x272
Attachment 8:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown