19:51 Shift change
Stats ok
ucesb ok, Max implants ~1500 Hz
20:30
Stats ok
ucesb ok, Max implants ~1000 Hz
current ok (maybe leveling out)
20:51
The online people are starting a new sort and are going to change some things with degraders I think. We are on R1_340.
20:59
Check
Clock status test result: Passed 16, Failed 0
Calibration test result: Passed 16, Failed 0
WR:
Base Current Difference
aida05 fault 0x36ca : 0x36ce : 4
White Rabbit error counter test result: Passed 15, Failed 1
FPGA time:
Base Current Difference
aida13 fault 0xa : 0xb6 : 172
FPGA Timestamp error counter test result: Passed 15, Failed 1
--- both errors above same module as earlier shift
Returned 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0 0
Mem(KB) : 4 8 16 32 64 128 256 512 1k 2k 4k
aida01 : 1 3 2 0 1 3 3 2 3 3 6 : 36092
aida02 : 6 4 4 2 2 3 1 3 3 3 6 : 36280
aida03 : 6 3 2 0 0 3 3 2 3 3 6 : 36048
aida04 : 10 3 1 1 3 3 1 3 3 3 6 : 36272
aida05 : 19 8 2 0 2 3 1 3 3 3 6 : 36268
aida06 : 2 2 0 2 3 4 2 2 3 3 6 : 36120
aida07 : 6 7 3 0 3 4 3 2 3 3 6 : 36416
aida08 : 7 5 4 0 3 2 2 3 3 3 6 : 36420
aida09 : 21 6 3 2 2 4 2 3 2 3 6 : 35700
aida10 : 13 3 1 1 3 3 1 4 2 3 6 : 35772
aida11 : 25 6 4 2 1 4 2 4 2 3 6 : 36180
aida12 : 0 4 0 3 2 3 3 3 2 3 6 : 35712
aida13 : 18 5 2 3 2 3 2 4 2 3 6 : 36080
aida14 : 2 1 1 1 2 3 2 4 2 3 6 : 35904
aida15 : 0 1 1 3 0 4 2 2 3 3 6 : 35960
aida16 : 4 5 3 2 3 4 1 3 2 3 6 : 35432
ucesb ok elog:309/1
current ok elog:309/2
Stats ok elog:309/3
temps ok elog:309/4
21:36
ucesb ok
Stats ok
22:01
ucesb ok 1500 Hz peak implant
Stats ok
temps ok
bias ok - earlier rise seems to have stopped, 0.2uA fluctuations on values.
22:20 Nic noted that they were seeing fewer y-strips events (x agreed with FRS rate).
Dead time was suggested as a problem. Oscar investigating, good event rate in midas doesn't show significant dead time.
Looking at the difference in FEEs3 & 4, which should have the same scaler rate but don't, suggestion data lost in the pause and resume.
Oscar is going to raise the threshold on one of the y FEEs as a test
Increased aida4 slow comparator threshold changed
There is some confusion as it seems to be dead time but apart rates shown look low enough, dead time is possibly masking true rate?
Results of test, need thresholds 100keV on p+n and 320keV n+n to get a roughly 1:1 between FEE3 and FEE4 scaler
23:40 FRS magnet issue, possibly power, no beam
00:03 bias ok elog:309/5 has come down because no beam
handing over to muneerah |