Message ID: 380     Entry time: Fri Sep 3 13:39:42 2021
Author: OH 
Subject: Additional noise tests performed at GSI 
Contents of an email from NH to OH and TD about tests performed on an additional bias supply:

This email was sent to you by someone outside the University.
You should only click on links or attachments if you are certain that the email is genuine and the content is safe.

so I looked at the other supply in the lab (in a NIM crate there, cable direct to DSO)

Still see the same noise wave - also it looked a lot worse at 50V but that may be a worse NIM crate or something else - the current also wasn't 0 (calibration?)

More unusually the signals remain even if the NIM crate is off (DSO_Lab_NoP) and when the SHV is removed from the HV completely (DSO_Lab_NoCable)
Attachment 1: DSO_Lab_NoP.jpg  170 kB  Uploaded Fri Sep 3 14:41:17 2021  | Hide | Hide all
DSO_Lab_NoP.jpg
Attachment 2: DSO_Lab_1.jpg  202 kB  Uploaded Fri Sep 3 14:41:17 2021  | Hide | Hide all
DSO_Lab_1.jpg
Attachment 3: DSO_Lab_50V.jpg  177 kB  Uploaded Fri Sep 3 14:41:17 2021  | Hide | Hide all
DSO_Lab_50V.jpg
Attachment 4: DSO_Lab_NoCable.jpg  17 kB  Uploaded Fri Sep 3 14:41:17 2021  | Hide | Hide all
DSO_Lab_NoCable.jpg