Message ID: 388
Entry time: Thu Oct 21 16:18:51 2021
Reply to this: 389
|
Author: |
OH |
Subject: |
FEE and Adaptor board groundings |
|
|
During investigations into the noise today by NH and OH a couple of findings were made on the grounding of the FEEs and Adaptor boards.
- The adaptor boards have no connection to the FEE ground over the ERNI connector
- Currently we are providing one via a ground cable between the FEE Al cooling and the Lemo ground on the adaptor board
- The Al cooling plate is not electrically isolated from the copper plate on the ASIC and both are connected to the electrical ground on the FEE
- Verified with 0 resistance between plate and HDMI ground shield
- The FEE is electrically isolated from the frame but when cabling is connected they show 0 resistance
- With electrics disconnected but water still connected they show 200kOhm
- With the bias core going to FEE14 and the braid going to FEE8 better performance in FEE14 is obtained by disconnecting FEE8 adaptor board from the ground
- The cost of this is the rate of FEE8 goes up to 300k
- It is essential to have the adaptor board ground on the FEE ground but unsure why having both adaptor boards connected to their respective grounds makes one worse
- The resistance between all FEEs is not zero. Some show a 5Ohm and some show a 35Ohm resistance between each other
|