> > Attachments 1 & 2 per FEE64 1.8.L spectra > pulser peak widths aida01 121 ch FWHM, aida02 517 ch FWHM > > Attachments 3 & 4 per FEE64 1.8.W spectra 20us FSR > > Attachment 5 per FEE64 rate spectra > > Attachment 6 - grafana DSSSD bias/leakage current for previous 7 days showing effect of day/night temp variations > > Attachments 7 & 8 - merger & tape server c. 14Mb/s to disk - no errors reported by servers > > Attachments 9-12 - system wide checks - all OK > > Attachment 13 ADC data item stats > > Attachment 14 FEE64 temps OK > > Attachment 15 DSSSD bias & leakage currents OK > > 16.15 aida05 zero ADC data items stats > merger restart, DAQ STOP/Go cycle to recover > > > activity is S4 area near AIDA DSSSDs as install of DEGAS detectors commences > > 19.50 DEGAS install continues > > DSSSD bias and FEE64 power OFF >