Entry time:
Thu Apr 3 19:29:14 2025
Author
*
:
Subject
*
:
Arrived at 09:54 - sequenced all FEE64 on FEE64 not mounted properly - something happened to network overnight, see attachment 1,2 for AIDA system log 10.23 - FEE64 still not mounted properly, power cycling FEE64 10.30 - sequenced all FEE64 on 10.43 - FEE64 mounted successfully - DAQ reset and setup - looks ok - histogram enabled on aida12 - clock ReSYNC - system wide checks - all FEE64 pass clock status check - aida2 and aida12 fail ADC calibration - SYNC pulse and SYNC error checks ok - WR decoder check returns errors (attachment 3) 10.53 - FEE64 temperatures ok (attachment 4) - detector bias/leakage current ok (attachment 5) 11.04 - TapeServer start - not writing to storage - ASIC check load performed - DAQ start - ASIC control tab (attachment 6) - Run control keeps reloading - has returned to 'reset' screen with aida11 showing "undefined" (attachment 7) - good event statistics - all FEE64 showing 0 *except* aida09 (attachment 8) 11.20 - stopped tape server - reset daq - returns error (attachment 9) 11.41 DAQ off, detector bias off, FEE64 off 13.30 detector bias on, FEE64 sequenced on all browser tabs reopened merger set back up - ok mbs data relay set back up - ok DAQ setup and started, initially ok good event statistics (attachment 10) 16.25 FEE64 restarted, along with Merger, TapeServer, DAQ (not writing to file) thresholds (slow comparator, fast comparator HEC, fast comparator LEC) raised to 0xff for testing. aida10 stats appear reasonable, however wave events still reported despite being deactivated (see report entry on DAQ stall) (attachment 11) 17.00 DAQ, TapeServer, Merger all stopped Detector bias off FEE64s switched off coolant water still running (cannot be disabled due to lack of access) AIDA in safe state
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown