12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no data, ASIC temperature c. 20 deg C ( ambient ) low Replaced ASIC mezzanine with new ASIC mezzanine Installed as aida08 Water temperature & pressure as measured outside S4 area - OK Manually power FEE64s BNC PB-5 local control/ON Amplitude 1.0V Attenuation x10 Polarity + tau_d 1ms Frequency 22Hz FEE64 temps - attachment 1 aida02 with new ASIC mezzanine continues to ramp to c. 512 deg C - initially reads 0, next refresh c. 70 and third refresh c. 512 deg C aida02 Virtex and PSU temps OK, aida02 cooling plate ambient to touch aida08 ASIC temp c. 45 deg C as expected, Virtex and PSU temps OK All system wide checks OK WR timestamp OK - attachment 2 aida04 ASIC settings - attachment 3 ADC data item stats OK - attachment 4 aida04 Rate spectrum - attachment 5 aida04 *.*.L spectra - attachments 6-9 1.8.L pulser peak width 13 ch FWHM aida04 1.8.W spectra - 20us FSR - attachment 10-11