Entry time:
Sat Mar 29 11:24:29 2025
Author
*
:
Subject
*
:
> 12.30 FEE64 41:d7:cd ASIC mezzanine u/s, no data, ASIC temperature c. 20 deg C ( ambient ) low > Replaced ASIC mezzanine with new ASIC mezzanine > Installed as aida08 > > Water temperature & pressure as measured outside S4 area - OK > > Manually power FEE64s > > BNC PB-5 local control/ON > Amplitude 1.0V > Attenuation x10 > Polarity + > tau_d 1ms > Frequency 22Hz > > FEE64 temps - attachment 1 > aida02 with new ASIC mezzanine continues to ramp to c. 512 deg C - initially reads 0, next refresh c. 70 and third refresh c. 512 deg C > aida02 Virtex and PSU temps OK, aida02 cooling plate ambient to touch > > aida08 ASIC temp c. 45 deg C as expected, Virtex and PSU temps OK > > All system wide checks OK > > WR timestamp OK - attachment 2 > > aida04 ASIC settings - attachment 3 > > ADC data item stats OK - attachment 4 > > aida04 Rate spectrum - attachment 5 > > aida04 *.*.L spectra - attachments 6-9 > 1.8.L pulser peak width 13 ch FWHM > > aida04 1.8.W spectra - 20us FSR - attachment 10-11
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown