<p>new Downstream DSSD2: 3208-2/3208-5/3208-8<br />
<br />
Covered with black cloth.</p>
<table border="1" cellpadding="1" cellspacing="1" style="height:346px; width:490px">
<tbody>
<tr>
<td>Voltage (V)</td>
<td>Current (uA)</td>
</tr>
<tr>
<td>10</td>
<td>3.1</td>
</tr>
<tr>
<td>20</td>
<td>3.8</td>
</tr>
<tr>
<td>30</td>
<td> </td>
</tr>
<tr>
<td>40</td>
<td> </td>
</tr>
<tr>
<td>50</td>
<td>4.515</td>
</tr>
<tr>
<td>60</td>
<td> </td>
</tr>
<tr>
<td>70</td>
<td>4.2</td>
</tr>
<tr>
<td>80</td>
<td>4.545</td>
</tr>
<tr>
<td>90</td>
<td>5.5</td>
</tr>
<tr>
<td>100</td>
<td>7.940</td>
</tr>
<tr>
<td>110</td>
<td> </td>
</tr>
<tr>
<td>120</td>
<td> </td>
</tr>
</tbody>
</table>
<table border="1" cellpadding="1" cellspacing="1" style="width:500px">
<thead>
<tr>
<th scope="col">
<p>Voltage (V)</p>
</th>
<th scope="col">Current (uA)</th>
</tr>
</thead>
<tbody>
<tr>
<td>10</td>
<td> </td>
</tr>
<tr>
<td>20</td>
<td> </td>
</tr>
<tr>
<td>30</td>
<td> </td>
</tr>
<tr>
<td>40</td>
<td> </td>
</tr>
<tr>
<td>50</td>
<td> </td>
</tr>
<tr>
<td>60</td>
<td> </td>
</tr>
<tr>
<td>70</td>
<td> </td>
</tr>
<tr>
<td>100</td>
<td>7.6</td>
</tr>
<tr>
<td>110</td>
<td>9.5</td>
</tr>
<tr>
<td>120</td>
<td>Breakdown</td>
</tr>
<tr>
<td> </td>
<td> </td>
</tr>
<tr>
<td> </td>
<td> </td>
</tr>
</tbody>
</table>
<p>Wafer 3 (beam left) shows breakdown issues at 90V... curious</p>
<p>We replace with november beam triple: 3208-3/3208-21/3208-22</p>
<p>We must speak to Micron about why so many 3208 wafers seem to have issues at >90V<br />
It seems unlikely to be other issues</p> |