AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  DESPEC  ELOG logo
Entry  Fri Jun 28 15:05:49 2019, NH, Aida 28/6/2019 - WATER OFF 
    Reply  Mon Jul 8 16:02:52 2019, NH, PJCS, Aida 28/6/2019 - WATER OFF 
Message ID: 56     Entry time: Mon Jul 8 16:02:52 2019     In reply to: 54
Author: NH, PJCS 
Subject: Aida 28/6/2019 - WATER OFF 

It occurs to to wonder if there is now water inside the modules. If the metal of the cooling plates is damp on the outside then it is safe to assume there is water on the inside too . Can I suggest that one  of the modules is dismounted, disassembled and checked?

Patrick

Quote:

The humidity sensor hasn't reported humdities under 90% yet. In order to try and help dry the pipes and turn AIDA the water has been turned off the weekend.
Sunday temperatures might reach 40 C which means it may struggle anyway.

Will consider trying to get a temperature/humidity monitor for S4 to check if the environment is fine too.

 

ELOG V3.1.4-unknown