> 11.45 DSSSD bias & leakage current OK - attachments 9-10 > > FEE64 temps OK - attachment 8 > *except* aida02 ASIC temp which is known to be u/s > > All system wide checks OK *except* aida02 and aida04 WR decoder/FPGA status errors - attachments 6-7 > HDMI cabling disturbed during platform move? > > WR timestamps OK - attachment 5 > > ADC data item stats - attachment 4 > 10/16 < 20k, max 115k > > per FEE64 Rate spectra - attachments 3 > > per 1.8.W spectra - 20us FSR - attachments 1-2 > > 11.58 BB7 preamps unplugged. Noise conditions OK. Did not change considerably. > 11/16 < 20k (aida09 has 1 hot channel), max 114k Attachments 11-12. > > 12:34 Detector powered down to bring to safe state to check HDMI cables of aida02 and aida03. > > 15:06 Reseated aida02 and aida03. No more FPGA timestamp error. See attachment 13-14. > > Rates and histograms look good for pulser walk through. Test + and test - connected. Currently, Attachment 16-17. > Temp OK Attachment 15. Bias OK Attachment 18. > > Data links aida02 and aida08 missing (?) > > 16:30 All slow comparator thresholds set to 0x64 for alpha background run during platform freeze. > > Initial rates and histograms attachment 19 - 20. aida04 noisy in ASIC 1. > > Temps OK. Attachment 21. > MBS disc and tape setup in atttachment 22-23.