<p>14:50 Checks</p>
<p>Leakage current is 20.021 and HV status is Max Voltages ! Is it normal ? - attachments 1</p>
<p>It is quite warm at GSI today and this may be a temperature effect. Hopefully will go down latter in the day (Marc). Any way let's kep an eye it. The current threshold is set to 30uA.</p>
<p>18:15 Checks</p>
<p>Everything appears to be going smoothly </p>
<p>Grafana - DSSSD bias and leakage current - attachments 2 and 3</p>
<p>FEE64 temperatures - It appears okay, nothing strange , attachment 4</p>
<p>ADC data item stats - attachment 5</p>
<p>Per FEE64 Rate spectra - attachment 6</p>
<p> ucesb - attachment 7</p>
<p>Merger Link Data Rates - attachment 8</p>
<p> </p>
<p>18:16</p>
<p>The temperature for aida07 gave us No response - attachment 9 , I had emailed Tom to fix.</p>
<p>18:20 it came back to work normally</p>
<p>............................</p>
<p>20:00 Checks</p>
<p>Nothing new to report. All is well.</p>
<p>20:33 Aida02 showed as not reading - attachment 10</p>
<p>20:34 it resumed reading and returned to normal operation.</p>
<p>21:08 Aida02 was not reading - attachment 11 , had to emailed Tom to fix it .</p>
<p>The shift crew have restarted ucesb and AIDA is now showing OK - attachment 12</p>
<p>so Tom guess the problem really was at their end . They may still have a problem with the FRS DAQ.</p>
<p>ADC data item stats and Merger etc appear to be going smoothly - attachments 13 and 14</p>
<p>...........................................</p>
<p>22:00 Checks</p>
<p>Still running smoothly.</p>
<p>DSSD bias and leakage current - ok</p>
<p>FEE64 temperatures - ok</p>
<p>Per FEE64 Rate spectra - ok</p>
<p>ADC data item stats - ok</p>
<p>Ucesb - ok</p>
<p>.................................</p>
<p>23:00 Checks</p>
<p> Nothing new to report. Everything is going well , and I have captured screenshots of each one</p>
<p>Leakage currents dropped a bit See attachment 15</p>
<p>.............................</p>
<p>23:50 Checks</p>
<p>All seems smooth.</p> |