Per NH
"From ELOG (assuming times accurateish)
10:00 Source in (AIDA removed from DESPEC DAQ)
11:30: Source out (?)
11:45: Background Runs
13:00: bPlast thresholds changed
15:45: Beam Returned"
-
ADC data item rate
EOF time file LEC HEC
09.02 R4_538 0.906M 1.7k
10.02 _551 0.927M 1.7k
11.00 _564 0.875M 1.7k
11.19 _568 0.872M 1.7k
12.18 R6_7 1.121M 1.9k
13.03 6_19 1.134M 1.9k
14.02 6_35 1.197M 2.0k
15.00 6_52 1.232M 2.0k
15.10 6_54 1.225M 2.0k
15.36 6_56 1.467M 2.0k
16.00 6_65 1.630M 3.0k
17.01 6_89 1.778M 3.7k
21.43 6_207 1.759M 2.7k
12.6.24
00.09 6_274 2.258M 3.8k
03.12 6_372 2.295M 4.4k
07.13 6_506 2.284M 4.5k
11.12 6_611 1.853M 3.6k
14.55 7_16 1.877M 4.7k
13.6.24
07.01 7_457 1.788M 4.3k
12.34 7_610 2.000M 4.7k
14.00 7_649 1.756M 2.5k
14.12 7_654 1.817M 2.5k
15.33 7_693 1.907M 3.5k
14.6.24
07.23 8_35 1.677M 3.7k
10.54 8_122 1.880M 4.7k
14.38 8_213 1.935M 3.6k
15.6.24
01.25 9_27 0.951k 1.6k (beam off from c. 18.15 14.6.24)
08.56 9_152 1.215M 2.1k gamma source in use?
15.27 9_275 1.520M 2.3k 370kBq 152Eu source in use
16.6.24
15.06 9_747 1.667M 2.3k
15.06 analysis data file R9_747 - attachment 18
max dead time 4.3% aida08, all other FEE64s < 2%
ADC data rate 1.667M, HEC data rate 2.3k |