<p>Yesterday we spent time essentially performing a dry run to get AIDA DAQ and bPlast into the time sorter, there were some issues with the AIDA mbs PC x86l-119 which was related to some boot issue, it was booting to a newer version of debian while the Relay for AIDA to MBS is on an older version (scratch) (the machine is quite old). We spent a bit of time getting the thresholds correct for each of the ASICs. In the end we just elected to have a blanket level of 0x0f for all of the ASICs when we removed the pulser.<br />
<br />
Last night we left the setup with the following thresholds on all of the FEEs to collect data. 681/1. The pulser was also turned off and the 22Na source was moved close to the snout.</p>
<p>This morning c. 9:25 we returned and checked the temperatures, HV and statistics and everything seems to be ok. The only problem was that the timesorter has crashed this morning at 8 am.</p> |