10.05 DSSSD # 1-3 detector biases & leakage currents OK - see attachment 1 Ambient temperature +21.2 deg C, d.p. +3.4 deg C, RH 31.2% 10.15 DSSSD stack as follows DSSSD serial thickness depletion # (um) voltage (V) 1 3208-7 1019 125 upstream 2 3208-13 1020 120 3 3208-14 1020 130 downstream 11.30 AIDA power relays sequenced on DAQ start system wide checks ok 11.33 attachment 2 - fee temperatures - ok attachment 3 - good event statistics with slow comparator threshold = 100 attachment 4 - good event statistics with slow comparator threshold = 50 attachment 5 - good event statistics with slow comparator threshold = 20 attachment 6 - good event statistics with slow comparator threshold = 10 13.51 attachment 7 - Rates - aida09 1 ASIC not working (no data) attachment 8 - good event statistics attachments 9/10 - 1.8.L spectra - no peak observed in aida09 attachments 11/12 - 1.8.W spectra - waveform only observed in naida 2,4,7,8,12 pulser peak widths - FEE width(ch) 1 71 2 96 3 66 4 79 5 74 6 96 7 89 8 83 9 no peak 10 82 11 153 12 120 17.26 FATIMA moved upstream around AIDA attachment 13 - good event statistics ASIC check-load performed attachment 14 - good event statistics following checkload