<p>Attachment1: AIDA05.1.8L when AIDA05 connected, biased at 101 V with shielded cables. BB7 is still connected but not biased (same configuration as the previous elog). The channel of the CAEN power supply #0 is connected to ASIC08 and biased at 100V. The channel of the CAEN power supply #1 is disconnected but still biased at 100V. AIDA04 (chained to AIDA05) is biased via a shielded cable to an iseg NHQ 246L power supplied, biased at 101 V with 16 uA.</p>
<p>All the outside drain wires of bplast are now taped to the snout.</p>
<p>The red markers of the cursors of the previous test have been left for ease of comparison.</p>
<p>FWHM: 21699-21644 = 55 channels (to be compared with the previous elogs between 73 channels and 79 channels)</p>
<p><strong>Conclusion</strong>: a significant improvement can be observed with a cleaner power supply.</p> |