AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
DESPEC
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Fri Nov 7 14:46:35 2025
Author
*
:
Subject
*
:
19.45 Remote system check DSSSD bias & leakage currents OK - attachment 1 FEE64 temperatures OK - attachment 2 aida02 ASIC temperature often reporting zero but sometimes a sensible value - presumably PT100 fault ADC data item stats - attachment 3 all FEE64s at high/very high rates - only 3x of 16x FEE64s < 100k ASIC settings per aida01-aida04 p+n FEE64s slow comparator 0xa, n+n FEE64s 0xf cf. 0x64 used for previous tests by Nabil? per FEE64 Rate spectra - attachment 4 per FEE64 1.8.W spectra - 20us FSR - attachments 5-6 19.55 Switch to safe state (FEE64 power OFF, DSSSD HV OFF) overnight as appears to be unattended/not monitored overnight. Tomorrow I would suggest resuming tests (using slow comparator 0x64) and checking per FEE64 pulser peak width for comparison with previous tests by Nabil. If unable to reproduce previous tests (or previous AIDA performance) you will need to check setup/reduce noise to be able to perform measurements using the 207Bi source.
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Attachment 1:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.3-7933898