| <p>Some FEES were not acquiring data for some reason we could not understand. We lowered the thresholds to 200 keV, then rose them again to 700 keV. It seems they are now acquiring. We start data taking again.</p>
<table align="center" border="1" cellspacing="1" style="border:1px solid #486090; width:98%">
<tbody>
<tr>
<td style="background-color:#486090">Quote:</td>
</tr>
<tr>
<td style="background-color:#ffffb0">
<p>- Bias and basic checks: the pulser FWHM appears to be ~170, which is the same we have obtained at the end of last week.</p>
<p>- Initialised the Merger and Tape server</p>
<p>- loading of settings: EXPERIMENTS/AIDA/2025Nov03-16.21.44</p>
<p>- slow comparator thresholds set to 700 keV for all X FEES</p>
<p>- y FEEs disabled in the discriminator</p>
<p>- Starting data taking with following details:</p>
<p>MIDAS data in: Bi207Centre_700</p>
<p>MBS data in /lustre/despec/aida_sourcetest_2025/raw started from lxg3138 on screen ('AIDAdaq')</p>
<p> </p>
<p>We take data overnight in this configuration and continue the grounding checks tomorrow</p>