AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
DESPEC
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Wed Nov 13 09:10:37 2019
Author
*
:
Subject
*
:
10:10 moved HV cables from 11 and 12 to 10 and 9 - all detectors biased in same configuration lowered slow comparator threshold to 0xa DAQ start pulser on zero'd histograms pulser peak widths: FEE width (ch) aida01 - 130 aida02 - 168 aida03 - 83 aida04 - 83 aida05 - 200 aida06 - 148 aida07 - 250 aida08 - 81 aida09 - 134 aida10 - 190 aida11 - 333 aida12 - 149 For comparison, peak widths from yesterday's measurement: pulser peak widths - FEE width(ch) 1 128 2 145 3 84 4 74 5 207 6 125 7 172 8 80 9 193 10 183 11 238 12 138 Attachments 1 & 2 : pulser peak spectra Attachment 3 - good event statistics Attachment 4 - detector bias/ leakage currents
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
191113_pulserpeaksOdd.png
Original size: 1920x1080
Attachment 2:
191113_pulserpeaksEven.png
Original size: 1920x1080
Attachment 3:
191113_goodevents.png
Original size: 1920x1080
Attachment 4:
191113_bias.png
Original size: 502x344
Attachment 5:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown