Entry  Thu Jan 9 16:25:51 2025, TD, LEC fast comparator threshold too low Screenshot_from_2025-01-09_17-28-45.png

LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
    Reply  Thu Jan 9 17:36:50 2025, TD, LEC fast comparator threshold too low 

> LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
    Reply  Thu Jan 9 17:56:14 2025, TD, LEC fast comparator threshold too low 
> > 
> > LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
> > 
    Reply  Fri Jan 10 10:03:43 2025, TD, LEC fast comparator threshold too low 
> > > 
> > > LEC fast comparator has been set to 0x2 (200keV) which is *very* low and means the fast comparator will be triggering on noise.
> > > 
Entry  Sat Jan 11 14:05:03 2025, TD, Saturday 11 January 2024 17x

14.37 Detector bias & leakage current OK - attachment 1
  
Entry  Sat Jan 11 14:14:21 2025, TD, aida06 crashed and rebooted 10.1.25 
aida06 crashed and rebooted sometime ( 14:48 - 23:25 ) yesterday. System console log appended.

Entry  Mon Jan 13 11:32:30 2025, TD, Offline analysis S505 data file R3_150 22x
Offline analysis of S505 data files R3_150

first WR ts
Entry  Mon Jan 13 17:38:18 2025, TD, Offline analysis of S505 data files 10x

First timestamp of R2_105 0x16FAEC5E54AABF6C
Entry  Wed Jan 15 11:58:08 2025, TD, Offline analysis data files S505 R3_9 - R3_128 (207Hg setting) 27x

Aanlysis data files R3_9 - R3_128 (207Hg setting)
Entry  Wed Feb 12 10:46:01 2025, TD, Offline analysis data files S181 R4_351-396  25x
Analysis data files R4_351-396

Entry  Sat Feb 22 12:35:38 2025, TD, Offline analysis data files R9_85-R9_199 (84Mo setting) 22x

Start 00:43 22.2.25 https://elog.gsi.de/despec/G-24-00302/57
AIDA data file R9_85
Entry  Tue Mar 11 11:53:18 2025, TD, Monday 10 March 11x

18.30 AIDA powered up for brief test
Entry  Tue Apr 8 08:52:39 2025, TD, AIDA grounding powerstax-ms1u_instructionmanua.pdf20250408_095624.jpg

AIDA DSSSDs and FEE64-DSSSD cabling are electrically isolated (by design) from snout, mounting rods and support assembly.
 FEE64-DSSSD ribbon cables screened by 3M 1245 Cu foil, drain wire to FEE64 adaptor PCB
Entry  Sat Jun 14 11:20:45 2025, TD, Saturday 14 June 2025 30x

12.21 BNC PB-5 settings
      Amplitude 1.0V
Entry  Sun Jun 15 10:53:08 2025, TD, Sunday 15 June 2025 41x

11.45 DSSSD bias & leakage current - attachment 1
      FEE64 temps OK - attachment 2
Entry  Thu Oct 23 18:45:12 2025, TD, Thursday 23 October 2025 6x

19.45 Remote system check
Entry  Sun Oct 26 16:22:38 2025, TD, [How To] Check ASIC settings 

tcsh
cd /MIDAS/DB/EXPERIMENTS/AIDA/<ASIC settings DB key>
    Reply  Mon Oct 27 18:08:11 2025, TD, Changing ground setup from 25102025 
19.06 Found HV OFF, DAQ GOing

      DAQ STOP
    Reply  Wed Oct 29 12:56:29 2025, TD, Further setup for 207Bi test 
Remember that 'ADC calibration' refers to the setup of the fast serial output of the AD9222 Octal, 50MSPS, 14 bit ADC which generates the waveforms
- if the FEE64 is unable to calibrate all that happens is that you will not be able to see preamp output waveforms. Energy spectra will work regardless
as they use a different ADC.
Entry  Wed Oct 29 13:39:13 2025, TD, AIDA FEE64 adaptor PCB to AIDA DSSSD cabling 

The cables form the AIDA FEE64 adaptor PCB to the MSL type BB18 DSSSDs are constructed as follows