<p>Start: 16:50</p>
<p>Stop: 17:28</p>
<p>Root: 201906071728_60Co_AIDA_in.root</p>
<p>Comment: 60Co source on top of AIDA (red box, key 3). It is positioned on a plastic ruler in the center of the DTAS.</p>
<p>File Number Aida: R18 (_0 should not be used, reset of the time stamp was done during this file). R18_1 started at 16:57, last file R18_4</p>
<p>File Number DTAS: 245,247</p>
<p>201906071650_60Co_AIDA_245.dlt,... 201906071650_60Co_AIDA_247.dlt</p>
<p>After checking the shifts in gain, it seems that for most of the detectors the dynode and anode signal moved in the same direction with the exception of detector 13 (dynode and anode do not move</p>
<p>equally, so probably there is a cable problem here, probably in the Dynode 13 or preamp problem)</p>
<p>Edit 13/06/19 Oscar Hall</p>
<p>The low energy channels of AIDA have been gain matched. Attached is the energy plot of the x and y side strops in AIDA.</p> |