AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
DTAS
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Mon Jun 10 04:18:24 2019
Author
*
:
Subject
*
:
File: 20190610_1145_Conf_DTAS.xlsx Several changes and corrections Naming: V2A8C1 -> Aux01, ... New or corrected Aux13 -> LP_out (puse out of light pulser) Aux15 -> DTpulser (Out of GSI clock triggering LP) Aux16 -> SyncPulser (Out of GSI clock send out for synchronization with AIDA & BigRIPS) Ranges changed Aux09, ERefAn We observed that LP_out was counting less than the other two so we tried to check DACQ parameters. We basically change nothing Only activated "CFD time" Note that pulse out is a very strange signal: Negative(?),-400mV offset, 200mV high In the DACQ is treated as positive signal with a large negative offset.
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
20190610_1145_Conf_DTAS.xlsx
Attachment 2:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown