AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
DTAS
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Fri Mar 28 20:50:31 2025
Author
*
:
Subject
*
:
> File: 20190610_1145_Conf_DTAS.xlsx > > Several changes and corrections > Naming: V2A8C1 -> Aux01, ... > > New or corrected > Aux13 -> LP_out (puse out of light pulser) > Aux15 -> DTpulser (Out of GSI clock triggering LP) > Aux16 -> SyncPulser (Out of GSI clock send out for synchronization with AIDA & BigRIPS) > > Ranges changed > Aux09, ERefAn > > We observed that LP_out was counting less than the other two > so we tried to check DACQ parameters. We basically change nothing > Only activated "CFD time" > Note that pulse out is a very strange signal: > Negative(?),-400mV offset, 200mV high > In the DACQ is treated as positive signal with a large negative offset.