Message ID: 131     Entry time: Sun Jun 16 07:08:10 2019
Author: Ela, Sonja, Victor, Alejandro 
Subject: problems in coincidence crystals-F11 
We have noticed that there is a problems in the coincidence Crystals-F11. The TAC shows a lot of counts. The
Crystals spectrum in coincidence with F11 do not. The time window of this coincidence is from-50 to 50 samples
and it should be from -180 to 300 samples to be centered in the peak of the TAC.

We will change it at soon as possible
Attachment 1: Crystals_coincidence_F11.png  13 kB  Uploaded Sun Jun 16 08:10:24 2019  | Hide | Hide all
Crystals_coincidence_F11.png
Attachment 2: TAC_crystals_coincidence_F11.png  20 kB  Uploaded Sun Jun 16 08:15:25 2019  | Hide | Hide all
TAC_crystals_coincidence_F11.png