<p>Short file with the strong Co source and Cs source. To test stability<br /> Start:11:32</p> <p>Stop: 14:04</p> <p>Root file: 201905201404_137Cs60Co_calib.root</p> <p>Rates; 53 kHz (Dynodes), 47 kHz (anodes) So thresholds should be checked</p> <p>Anode 12 (calibrated) has a large jump, because the cable in the digitizer was not fully plugged.</p> <p>We need to recalibrate them. The alignment was done originally with Dynodes.</p>