AIDA
GELINA
BRIKEN
nToF
CRIB
ISOLDE
CIRCE
nTOFCapture
DESPEC
DTAS
EDI_PSA
179Ta
CARME
StellarModelling
DCF
K40
DTAS
Draft saved at 00:00:00
Fields marked with
*
are required
Entry time:
Wed May 22 06:31:54 2019
Author
*
:
Subject
*
:
<p>Start: 14:32</p> <p>Stop: 14:37</p> <p>Source: 137Cs</p> <p>Comments: Detectors show small drifts, of the order of the precision of the HV (we could apply 0.5 V corrections to align some of them, but I will not touch them now)</p> <p>Dy16 looks quite off in the background measurement, but not with the Cs one !!!.</p> <p>Calibrated spectra: Dy3 an An3 moved a little in the same direction, An12 looks also a little displaced.</p> <p>Now I plan to measure some sources with AIDA in. It means that the sources will not be centered. The available distance from the oposite side to</p> <p>the nose of AIDA is 16 cm.</p>
Encoding
:
HTML
ELCode
plain
Suppress Email notification
Resubmit as new entry
Attachment 1:
201905221437_137Cs.root
Attachment 2:
Drop attachments here...
Draft saved at 00:00:00
ELOG V3.1.4-unknown