Start: 10:27
Stop: 11:31
Root: 201905251131_60co.root
Lmd: 60Co_source_201905251027_176.dlt ...60Co_source_201905251027_181.dlt
Comments: possitioned as the measurements of yesterday. 5 mm deeper than earlier.
Red box with the other sources inside BELEN, far away from the DTAS
A few detectors have moved, but for the moment I will not correct it.
Reference detector in channel 10. It is very noisy now. I changed the threshold from 1400 to 1600 but still noisy. It requires further attention
New Config file saved. |