<p>F11 plastic has been connected to our dacq as it was in the earlier BRIKEN setting.</p> <p>The detector cables were followed and connections checked.</p> <p>The outputs of the detector (L and R) go through an Ortec TFA 474 with the following settings:</p> <p>Coarse gain: x4</p> <p>Fine gain: looks minimal</p> <p>Integrate (ns): 200</p> <p>Diff (ns): 100</p> <p>Non-inv setting</p> <p>After the TFA, the Right signal is plugged in channel 1 of our card, and an attenuated signal (50 Ohm) is</p> <p>plugged in channel 3. Similarly the Left signal after the TFA is plugged in channel 2 and the attenuated </p> <p>signal is plugged in channel 4.</p> <p><br /> Parameters of the channels of the dacq have to be checked.</p>