<p>The position of the reference detector in the ancilliary card was changed from channel 10 to channel 9. Remember that this is important</p>
<p>in relation to all files saved until now.</p>
<p>The HV of AIDA plastic was changed from 1300 V back to 1000 V.</p>
<p>A new configuration file provided by Alvaro and Jose, with some small changes will be loaded.</p>
<p>cp 20190529_Conf_ATJLT.xlsx 20190530_Conf.xlsx</p>
<p>The new configuration file has been uploaded. Reference detector appears in channel V2A8C9. It was necessary to change the range of the histogram, and to recalibrate it.</p>
<p>With the new configuration parameters the pulser appears at 1360 keV in the calibrated spectra of the reference detector.</p>
<p>The threshold of the Ref detector has also been changed. From 4400 to 2800.</p> |