  100   Fri Jun 7 02:50:40 2019 SN, AAtest of 24Na

Start: 10:49

Stopt: 11:04

Lmd: 201906071049_24Na_AIDA_242.dlt

Root: 201906071117_24Na_AIDA_in.root

Comment: After irradiation of approximately 7 days, we extracted one Alu plate. It shows 24Na activity and also probably some beta+ activity.

The sample was not irradiated in the last 10 hours or so. It is positioned roughly in the center of the detector side in front of AIDA, using the frame prepared in Valencia.

It is not in the geometrical centre of the DTAS

Rates: 4.5 kHz (Dynodes), 4.5 (Anodes)  (Background level was at approximately 1.6-1.7 in both Dynodes, Anodes)

We need to check the alignment properly with Cs, but it looks that after one week some of the detectors have moved.

Group 1,3,5,7  (all seem to move with respect to detector 1)

Group 2,4,6,8 (detector 6 moved)

Group 11,13,15,17 (11 and 1 seem aligned, detector 13 moved)

Group 12, 14,16, 18 (12 and 2 not aligned, they seem aligned)

 

Attachment 1: 201906071117_24Na_AIDA_in.root
Attachment 2: Screenshot_from_2019-06-07_11-22-06.png
Screenshot_from_2019-06-07_11-22-06.png
Attachment 3: Screenshot_from_2019-06-07_11-24-39.png
Screenshot_from_2019-06-07_11-24-39.png
Attachment 4: Screenshot_from_2019-06-07_11-27-49.png
Screenshot_from_2019-06-07_11-27-49.png
Attachment 5: Screenshot_from_2019-06-07_11-30-17.png
Screenshot_from_2019-06-07_11-30-17.png
Attachment 6: Screenshot_from_2019-06-07_11-35-31.png
Screenshot_from_2019-06-07_11-35-31.png
Attachment 7: Screenshot_from_2019-06-07_11-36-43.png
Screenshot_from_2019-06-07_11-36-43.png
  12   Thu May 16 02:57:15 2019 Nacher, Algorathreshold dynode #13 increased
Threshold of dynode #13 changed from 500 to 700 because of a very high noise.
New configuration file: 20190516_Conf.xlsx (attached).
Attachment 1: 20190516_Conf.xlsx