Message ID: 12     Entry time: Sun May 23 14:02:58 2021
Author: CB, PB, LdF 
Subject: 2021 Runs - grade B MSL Si 
R1: 
Same as previous configuration? 
Negative signals.
No pulser
2048 window / 1000 pre-gate trigger
20 threshold / 44098 DC offset
3-alpha source

R2: 
Detector rotated. 
Negative signals
No pulser
2048 window / 1000 pre-gate trigger
20 threshold / 44098 DC offset
3-alpha source




R3: 
Detector rotated back. 
Positive signals.
2.5 Hz Pulser
1024 window / 700? pre-gate trigger
50 threshold / 48098 DC offset
3-alpha source


R4: 
Detector as before
Positive signals.
2.5 Hz Pulser
512 window / 600 pre-gate trigger
50 threshold / 48098 DC offset
3-alpha source

R5: 
Detector as before
Positive signals.
2.5 Hz Pulser
1024 window / 650 pre-gate trigger
50 threshold / 48098 DC offset
3-alpha source

R6:
As above. No alpha source.
Data acquisition stopped very early into the run. Lost contact with digitiser?


Upon re-establishing contact, signal digitisation looks extremely coarse in y.
Previously observed behaviour. Reasons unclear.
Calibrate ADC on V1730PSD panel seems to have fixed this. 
However, baseline/offset/threshold clearly changed based on data rate and signal.
Lowered threshold.

R7: 
Detector as before
Positive signals.
2.5 Hz Pulser
1024 window / 650 pre-gate trigger
40 threshold / 48098 DC offset
No source.
Data acquisition stopped very early into the run. Lost contact with digitiser again?

Did not reboot either PC or VME crate. Rebooted MIDAS only. Re-established contact. All looks good

R8:
As R7
Data acquisition stopped a few hours into the run. Same as before!

Did not reboot either PC or VME crate. Rebooted MIDAS only. Re-established contact. All looks good

R9:
As R7,R8
DAQ stopped again. Rebooted. OK. Issue needs to be understood.

R10
As above
DAQ did *not* crash!

R11
As above.

R12
As above, 208Bi source placed in chamber.
No obvious signal. Run stopped

R13
Pulser settings changed during optimisation.
Source location changed during optimisation.

---

Attempted to increase DC offset (->55098) and lower baseline (2400) to increase gain and make use of full
dynamic range.
Systems briefly triggers, contact with VME lost. Will need to reboot locally.