AIDA GELINA BRIKEN nToF CRIB ISOLDE CIRCE nTOFCapture DESPEC DTAS EDI_PSA 179Ta CARME StellarModelling DCF K40
  EDI_PSA  ELOG logo
Message ID: 3     Entry time: Mon Jun 10 06:17:51 2019
Author: CB, RG, AK 
Subject: 3-6 June - first DAQ tests 
Tested performance of DAQ with following equipment

Hamamatsu photodiode (S3590-09)
    - Powered by Farnell power supply (E30/1). Bias 14V / leakage current < 10uA.
Cooknell preamplifier
    - Powered by Farnell power supply (L30BT) @ 24V.

Self-triggering of the CAEN modules is not yet working. Currently triggering using analogue electronic chain.

Preamp signal is split. One half is sent to Ortec 747 TFA ( coarse gain = 1x / fine gain = 2 / int = 50ns / diff =
500ns / non-inv input) -> Octal CFD CF8000 ( threshold ?? ) -> LeCroy 688AL level adapter (for NIM to TTL
conversion) -> CAEN v1730 external trigger IN
Other half is delayed by 250ns line and sent to CAEN V1730 channel 0

Data saved in DiodeTes/
R1 -> Forgot bias to the diode
R2 -> Bias on. Overnight run
R3 -> Bias on. Short run to commission analysis program
ELOG V3.1.4-unknown